VT

คุณสมบัติสินค้า:

แบรนด์ : TOKYO KEISO

Share

Features

  • Simple installation
    • Wafer connection makes installation simple. Flowmeter body flanges are designed to match the pipe flanges by guiding the flowmeter in the center line by fastening bolts.
  • Short straight runs
    • The required straight runs are less than 1/5 of those required for orifi ce and vortex flowmeters.
      The narrow installation space allows simple and flexible piping arrangement plan.
      It leads to space and cost saving.
  • Low pressure loss
    • A proper selection of ß ratio allows lower pressure loss than orifi ce plate with the same flow rate. It improves energy efficiency of the plant.
  • 3-way stopcock installed
    • By operating its handle a newly developed 3-way stopcock can switch isolation, equalizing or measuring mode.
      This stopcock prevents mechanically such wrong operations as running off seal liquid in the impulse piping and overpressurizing to the one side of differential transmitter which might occur with the traditional three way manifold.
  • Wide variety of transmitters
    • Explosionproof type, pressure and temperature compensation type, and other types are available.
      You can make use of your transmitters also.

 

Short Specifications

Measuring objectLiquids, Gases, Saturated steam
Flow Range
Water: 
Max. 0 to 119.73 m³/h (100A, at 20°C)
Min. 0 to 5.51 m³/h (25A, at 20°C)
Air: 
ax. 0 to 10385 m³/h(nor) (100A, at 0°C,0.99MPa)
Min. 0 to 77m³/h(nor) (25A, at 0°C,0MPa)
Saturated steam: 
Max. 0 to 10745 kg/h (100A, at1.6MPA)
Min. 0 to 103 kg/h (25A, at0.1MPA)
Guaranteed
Accuracy Range
Max. Rangeability: 1:14
Fluid PressureLiquids: -98kPa to 2MPa
Gases: -98kPa to 1MPa
Saturated steam: -98kPa to 1.6MPa
Fluid TemperatureLiquids: -20°C to 120°C
Gases: -20°C to 120°C
Saturated steam: Max.204.3 °C
ConnectionWafer
Connection RatingJIS10K/20K,ANSI Class150/300,DIN PN16/40,GB PN1.6/4.0
Connection
(Meter) Size
25, 40, 50, 65, 80, 100(mm)
1, 1-1/2, 2, 2-1/2, 3, 4(inch)
Materials
Detector: 
Body: SCS14A
Cone: SUS316
H-Connector: SUS316/SCS14
Oval flange gasket: PTFE
3Way cock piece: 
Body: SCS14A
O-ring: Gasket PTFE and Fluorocarbon rubber Ethylene propylene rubber is also available.
Differential transmitter: 
It depends on a model
Series
/Power supply
/Output
VTG Series: 
Differential transmitter is provided by others.
VTF Series: 
FCX-AIII (Model FKC) differential pressure transmitter made by Fuji Electric Co. is installed in this series.
16.1 to 45 VDC / 2wire 4-20mADC(Non explosionproof or Flameproof by TIIS)
VTM Series: 
EDR-N7C differential pressure transmitter made by Hitachi High-Tech Control Systems Co. is installed in this series.
11.4 to 42 VDC / 2wire 4-20mADC(Non explosionproof or Flameproof by TIIS)
This instrument is used for the mass flow measurement of gas and saturated steam.
Accuracy±1.0 to 1.5 %(F.S.)


สินค้าเกี่ยวข้อง

Powered by MakeWebEasy.com
เว็บไซต์นี้มีการใช้งานคุกกี้ เพื่อเพิ่มประสิทธิภาพและประสบการณ์ที่ดีในการใช้งานเว็บไซต์ของท่าน ท่านสามารถอ่านรายละเอียดเพิ่มเติมได้ที่ นโยบายความเป็นส่วนตัว  และ  นโยบายคุกกี้