O3000

คุณสมบัติสินค้า:

แบรนด์ : TOKYO KEISO

Share

Features

  • Cost effective
    • The measuring method using an orifice plate enables the measurement of flow rates in medium- to large-size main pipes at low cost.
  • Omni-direction support (except O3□□□- AP)
    • Supports all directions from bottom to top, top to bottom, left to right, and right to left.
  • HART® communication
  • Explosion-proof
    • Certified by TIIS, KOSHA, NEPSI, ATEX and IECEx.
  • Indicator protection category: IP67

 

Short Specifications

Fluidliquids (up to viscosity 3 mPa·s)
gases in general (not suitable for slurry and steam)
Flow range
water (density 1.0g/cm³, viscosity 1.0 mPa·s)
Max.: 64 to 1000 m³/h (Diff. pressure: 50 kPa)
Min.: 0.27 to 1.8 m³/h (Diff. pressure: 30 kPa)
air (0℃, 1 atm)
Max.: 1700 to 26000 m³/h(nor) (Diff. pressure: 50 kPa)
Min.: 5.5 to 53 m³/h(nor) (Diff. pressure: 40 kPa)
Accuracy±5%F.S.
Temperature−20℃ to 120℃ (general type)
−20℃ to 200℃ (high-temperature type)
Pressureup to 2 MPa
Process connection
Screw connection: 
Rc tapered female thread and NPT female thread,
pipe size: 15 A (1/2") to 100 A (4")
Flange connection: 
JIS5KFF/10KFF·RF/20KRF,
ANSI/ASME/JPI class:150/300RF,
pipe size: 15 A (1/2") to 300 A (12")
Wafer connection: 
JIS5K/10K/20K,
ANSI/ASME/JPI class: 150/300,
pipe size: 15 A (1/2") to 300 A (12")
Tube material
Screw connection: 
SCS14
Flange connection: 
SUS304、SUS316、SS400/SGP
Wafer connection: 
SCS14、SUS304、SUS316、SS400
ConstructionNon-explosion-proof
Flameproof
Intrinsic safety
Output4 to 20 mA DC (electric transmitter, HART® communication)